More About Nebulizer, Model 503, Part #17068

NEBULIZER, MODEL 503, PART #17068 operates by electrically activating piezoelectric ceramic actuator (PZT) which then transduces the vibration generated to the adjacent supporting plate and polymer mesh bearing numerous apertures. The vibration actively pushes out the liquid medication by physically breaking surface tension of the solution through mesh holes thereby achieving final nebulization. Aerosol generation can subsequently be modulated the vibration frequency that is controlled by the firmware stored within Main Unit to maintain a frequency of approximate 117 kHz, nebulization rate of ≥0.25 ml/min and a power consumption of <1.5 W.

About HOSPAL MEDICAL CORP.

With over 30 years of experience in respiratory therapy, Hospal is dedicated to developing advanced nebulizer systems and solutions to improve the treatment experience for patients with respiratory diseases.